Agilent Cary 50 UV-Vis Spectrophotometer Guaranteed Specifications Manual

This document is a specification sheet for Varian Cary® 50 UV-Vis spectrophotometer. It mainly introduces the design overview, spectral range, wavelength range, resolution, spectral accuracy and other characteristics of the product.


Agilent Cary 50 UV-Vis Spectrophotometer Guaranteed Specifications Manual - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
80 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
1F2674B470E13AFDC6498E17B5FE1F3A

About this manual

The Agilent Cary 50 is a dual beam spectrophotometer featuring a Czerny-Turner monochromator with a wavelength range of 190 to 1100 nm. It utilizes a Xenon pulse lamp source and two silicon diode detectors, offering scan rates up to 24,000 nm/min and room light immunity.

What's inside

  • Design overview
  • Monochromator
  • Detectors
  • Wavelength range
  • Photometric accuracy
  • Photometric stability
  • Instrument dimensions
  • Environmental conditions
  • Electrical requirements
  • Warranty

Specifications

Spectral bandwidthFixed at 1.5 nm
Maximum scan rate24,000 nm/min
Photometric range (Abs)+ 3.3
Instrument weight21 kg
Compartment size130 mm x 623 mm x 123 mm
Minimum sample volumeApproximately 5.0 yL

Frequently asked questions

What is the warranty period for hardware support?

The hardware support period is twelve (12) months, though this may vary according to locations.

Can the sample compartment be left open during measurement?

Yes, the sample compartment can be left open due to the room light immunity of the Cary 50.

What is the photometric noise at 1 Abs?

The photometric noise is less than 0.0005 Abs (RMS) at 1 Abs.