Magnex MRBR 7.0/400 Magnet System Specification

This document is a specification for a 7.0 TESLA/400MM room temperature bore magnet system. The system is a complete superconducting magnet system intended primarily for research studies on the clinical/biological applications of NMR imaging (MRI) and NMR spectroscopy (MRS). The system consists of a highly homogeneous superconducting magnet (300MHz 1P, 7.05 Tesla) housed in a horizontal room temperature bore (400mm), low-loss helium cryostat. Field shimming is normally accomplished using both superconducting and room temperature shim coils. For imaging applications, X, Y, and Z gradient coils can also be provided with the system. Normally, the room temperature shims and gradients are mounted on independent non-conducting formers and are positioned.


Magnex MRBR 7.0/400 Magnet System Specification - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
15
File format
PDF
File size
483 KB
Published
04 July, 2012
Updated
28 September, 2023
MD5 checksum
0CFB6A9EBF34FFEB31C683FFC7A1A626

About this manual

This document provides technical specifications for the MRBR 7.0/400 system, a complete superconducting magnet system intended primarily for research studies on the clinical/biological applications of NMR Imaging (MRI) and NMR Spectroscopy (MRS). The system consists of a highly homogeneous superconducting magnet housed in a horizontal room temperature bore, low-loss helium cryostat.

The specification details the superconducting magnet, including its multi-filamentary NbTi conductor construction, field homogeneity, and operating current. It also covers the cryostat design, dimensions, and cryogen details for both liquid helium and liquid nitrogen. The document lists the scope of supply, including electronic consoles, gradient coils, and various monitoring units.

What's inside

  • Description of System
  • The Superconducting Magnet
  • The Cryostat
  • Scope of Supply
  • Superconducting Shim Coils
  • System Components

Specifications

Central field7.05 Tesla (300MHz 'P)
Field stabilityLess than 0.05 ppm/hour
Operating current178 Amps (nominal)
Field homogeneity (Superconducting only)Less than 20ppm over 20cm dsv
Field homogeneity (Fully shimmed)Less than 5ppm over 20cm
Room temperature clear bore400mm
Length of cryostat1840mm

Frequently asked questions

What is the primary intended use of the MRBR 7.0/400 system?

The system is intended primarily for research studies on the clinical and biological applications of NMR imaging and spectroscopy.

What is the central field strength of the superconducting magnet?

The magnet has a central field of 7.05 Tesla.