Agilent GC Troubleshooting Series Part Six: Retention Time Shifts

The document describes possible causes of retention time shifts in Agilent GC and provides solutions.


Agilent GC Troubleshooting Series Part Six: Retention Time Shifts - cover page
Brand
Agilent
Category
Other
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
93 KB
Published
04 July, 2012
Updated
01 October, 2023
MD5 checksum
AF78325294276B1C7A050AEF2220E793

About this manual

This document is part of the Agilent GC Troubleshooting Series, specifically addressing retention time shifts in gas chromatography. It outlines various potential causes for this issue, such as changes in column dimensions, leaks in the injector or gas lines, large changes in sample concentration, solvent incompatibility, inaccurate purge times, and column temperature variations.

What's inside

  • Changes in Column Dimensions
  • Leak in Injector or Blockage
  • Large Change in Sample Concentration
  • Solvent Incompatibility
  • Inaccurate Purge Time
  • Column Temperature
  • Leak Check Procedure
  • Manual Leak Check Steps

Frequently asked questions

What causes retention time shifts after changing a column?

It is probably due to changes in column length, as the old column may have been trimmed over time to a significantly shorter length.

What should I do if a leak check test fails?

Tighten the column fitting, septum nut, insert weldment nut, split vent trap housing, and gold seal fitting, then rerun the test.