This report describes an analysis method using GC/MS SIM technology to determine extractable amines in textiles and other materials, the method provides flexibility in low-level detection and can be used to analyze environmental samples associated with contamination monitoring of colorization process waste streams.
This document details the analysis of 20 carcinogenic amines banned in textiles and consumer products, as well as environmental monitoring of colorization process wastes. The study utilizes the Agilent 6890 gas chromatograph coupled with the 5973 mass selective detector (MSD). The system was found to be flexible enough to measure both high concentrations and trace amounts of the banned amines, meeting German regulatory limits.
| Detection Mode | SIM mode |
|---|---|
| Injection Method | Pulsed splitless injection |
The note analyzes 20 carcinogenic amines banned in textiles and consumer products.
The limit of detection (LOD) was determined to be 5 ppb using pulsed splitless injection and SIM mode.