Agilent 2100 Bioanalyzer Protein 200 LabChip Kit Absolute Quantitation Application Note

Absolute quantitation with the Agilent 2100 Bioanalyzer and the Protein 200 LabChip® kit. The Protein 200 LabChip® kit allows sizing and quantitation of proteins ranging in size from 14 to 200 kD. Relative quantitation is performed in comparison to the upper marker, used as internal standard in each sample. However, the accuracy of the relative quantitation is affected by the staining efficiency of the protein dye. Absolute quantitation is enabled with the Agilent 2100 bioanalyzer software (revision A.02.01) using the Protein 200 assay in combination with protein standards of known concentrations. Absolute quantitation improves quantitation accuracy by eliminating the differences in staining efficiency among proteins. This Application Note describes how to perform absolute quantitation with the Protein 200 assay and demonstrates the obtained improvement in accuracy.


Agilent 2100 Bioanalyzer Protein 200 LabChip Kit Absolute Quantitation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
405 KB
Published
03 July, 2012
Updated
28 September, 2023
MD5 checksum
03EB97ED47B397767A0C4E074C18009F

About this manual

The Protein 200 LabChip kit enables sizing and quantitation of proteins ranging from 14 to 200 kD. While relative quantitation uses the upper marker (myosin) as an internal standard, its accuracy is affected by varying staining efficiencies among different proteins. Absolute quantitation addresses this by using a calibration curve generated from standards of known concentration for the specific protein of interest.

What's inside

  • Application Abstract
  • Introduction
  • Results and Discussion
  • Conclusions
  • Calibration Curve
  • Relative vs Absolute Quantitation

Specifications

Protein Size Range14 to 200 kD
Software RevisionA.02.01
Relative Quantitation Reproducibilityapproximately ±30 %

Frequently asked questions

How does absolute quantitation improve accuracy?

It eliminates differences in staining efficiency among proteins by using a calibration curve generated with identical proteins of known concentrations.

What is used as the internal standard for relative quantitation?

The Protein 200 upper marker, which is myosin, is used as the internal standard in each sample.

Can BSA be used if standards of the target protein are unavailable?

Yes, BSA or another protein can be substituted to generate a calibration curve, though this results in relative quantitation rather than true absolute quantitation.