Agilent 1100 Series LC/MSD TOF Opiate Analysis Application Note

This article presents a method for quantitatively analyzing opiates in urine using liquid chromatography/mass selective detector time-of-flight (LC/MSD TOF). The method utilizes accurate mass measurement to enhance selectivity.


Agilent 1100 Series LC/MSD TOF Opiate Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
455 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
0BCB8226AA3EDF6B6B6065740F555E19

About this manual

This document presents an application demonstrating the quantitative analysis of opiates in urine samples using Agilent liquid chromatography/mass selective detector time-of-flight (LC/MSD TOF). The study highlights the advantages of accurate mass measurement for enhancing selectivity and reducing chemical noise.

What's inside

  • Abstract
  • Introduction
  • Sample preparation
  • Instrument
  • Experimental Conditions
  • Results
  • Conclusions

Specifications

InstrumentAgilent 1100 Series LC/MSD TOF
ColumnZORBAX XDB-C18, 2.1 mm x 50 mm, 3.5 pm
Mobile Phase AAcetonitrile with 0.1 % formic acid
Mobile Phase BWater with 0.1 % formic acid
Flow rate0.35 mL/min
Detection limit2 pg on-column

Frequently asked questions

What is the instrumental detection limit for this analysis?

The instrumental detection limit is 2 pg on-column with a signal/noise ratio of 5:1.

Which sample preparation methods are demonstrated?

The document demonstrates both direct injection of urine and solid phase extraction (SPE).

Download PDF (455 KB)