Agilent 5975 Inert MSD Chemical Ionization Technical Note

The 5975 inert MSD – Benefits of Enhancements in Chemical Ionization Operation Technical Note This Technical Note describes the complete integration of chemical ionization (CI) on the 5975 inert MSD (mass selective detector), which now makes all the power of CI mass spectrometry (MS) conveniently available. CI has been fully automated by providing auto-tuning with electronic control of the reagent gases. All modes of MS operation in a single sequence, completely free of operator intervention, are now possible. Analytes can be sensitively determined in samples by implementing alternating positive- or negative-CI (PCI or NCI) modes. On the 5975 inert MSD, CI is as easy as EI (electron impact ionization).


Agilent 5975 Inert MSD Chemical Ionization Technical Note - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
6
File format
PDF
File size
599 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
359D08CF2072198688779162991FF2E1

About this manual

The document details how the Agilent 5975 inert MSD integrates Chemical Ionization (CI) with Electron Impact (EI) ionization. Key enhancements include fully automated auto-tuning for both EI and CI modes, as well as complete electronic control of reagent gases (methane and ammonia). This integration allows analysts to switch between positive-ion chemical ionization (PCI) and negative-ion chemical ionization (NCI) within a single sequence without manual intervention.

What's inside

  • Cl Instrument Interface Enhancements
  • Automated EI tuning of CI source
  • PCI tuning with PFDTD
  • NCI tuning automation
  • Application Examples: Phthalates
  • Optimizing Source Temperature
  • Influence of Reagent Flow

Specifications

Reagent Gas OptionsMethane and Ammonia
Maximum MFC Flow5 mL/min

Frequently asked questions

What reagent gases are supported for CI on the 5975 inert MSD?

The instrument supports Methane and Ammonia as reagent gases.

Can PCI and NCI modes be run in a single sequence?

Yes, the instrument allows alternating between positive- or negative-Cl (PCI or NCI) modes within a single sequence.