Agilent G1888A Network Headspace Sampler Application Note for Residual Solvent Determination

The document introduces a method for determining residual solvents in pharmaceuticals using Agilent G1888 Network Headspace Sampler.


Agilent G1888A Network Headspace Sampler Application Note for Residual Solvent Determination - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
88 KB
Published
02 July, 2012
Updated
01 October, 2023
Model
G1888
MD5 checksum
A73220EC944D09F0CD8D697C6203E0D4

About this manual

The document describes an analytical method for determining Class 1 and Class 2 residual solvents in pharmaceutical products. It utilizes the Agilent G1888A Network Headspace Sampler coupled with a 6890N Gas Chromatograph configured with either a Flame Ionization Detector (FID) or a 5973 inert Mass Selective Detector (MSD). The method employs static headspace sampling to analyze solvents listed in USP 467 and International Conference on Harmonization (ICH) guidelines.

What's inside

  • Abstract
  • Introduction
  • Class 1 Solvents
  • Class 2 Solvents
  • Experimental Conditions
  • Instrument Settings
  • Calibration Curves
  • Headspace Equilibration

Specifications

Sample Loop Size1-mL
Split Ratio (Recommended)2:1 to 5:1
Detector Temperature (FID)260 °C
Headspace Oven Temperature85°C

Frequently asked questions

What solvents are covered in this application note?

The analysis covers Class 1 and Class 2 solvents according to ICH guidelines, as well as those listed in USP 467.

What is the recommended equilibration time for headspace sampling?

While normally set at 60 minutes, a 30-minute equilibration time is sufficient for most cases.

Download PDF (88 KB)

More Agilent Manuals


View all Agilent manuals →