Agilent 5973x GC/MSD Combined EI and CI Using a Single Source Application Note

This document introduces the optimized sources for electron ionization (EI) and chemical ionization (CI) of Agilent 5973x gas chromatograph/mass selective detectors (GC/MSDs). It also demonstrates how to acquire high-quality EI spectra with the CI source without changing the sources. The experimental results show that using the CI source can obtain comparable major and minor peaks to the EI source.


Agilent 5973x GC/MSD Combined EI and CI Using a Single Source Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
714 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3FFA80F998DF860F0CD374A5667CD574

About this manual

The document explains how to configure an Agilent 5973x gas chromatograph/mass selective detector equipped with a Chemical Ionization (CI) source to acquire Electron Ionization (EI) spectra. This capability allows users to obtain both EI data for library searching and CI data for molecular weight information without physically swapping ionization sources.

The text provides specific tuning steps, including reducing the methane reagent gas flow rate and adjusting the repeller voltage. It presents experimental results showing total ion chromatograms and mass spectra for a fragrance mixture (Hexadecanolide), demonstrating that the acquired EI spectrum matches the NIST02 library reference with 98% quality.

What's inside

  • Introduction
  • Data Acquisition
  • Tuning Procedure
  • Figure 1: Total Ion Chromatograms
  • Figure 2: CI Spectrum
  • Figure 3: EI Spectrum Comparison
  • Summary

Specifications

Document DateJanuary 30, 2004
Document Number5989-0595EN

Frequently asked questions

Can EI spectra be acquired using the CI source?

Yes, the Agilent 5973 inert MSD can acquire high-quality EI spectra with the CI source by adjusting settings like repeller voltage and gas flow.

What is the benefit of using a single source for EI and CI?

Acquiring both types of data without changing sources results in increased productivity.