Agilent HP-5MSi Column Application Note

The use of a highly inert capillary column can facilitate the quantitation of low levels of very active compounds. The use of the highly inert HP-5MSi column in conjunction with the Agilent 6890/5973 inert gas chromatograph/mass spectrometer (GC/MS) system allows the measurement of pesticides and environmental contaminants at low levels while maintaining response linearity over a wide concentration range.


Agilent HP-5MSi Column Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
184 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
550F4F93D596AC4A78706167EA97AC7A

About this manual

This document evaluates the performance of the new HP-5MSi (MS Inert) capillary column for the analysis of active compounds, specifically pesticides and environmental contaminants. The column features low bleed and is specially deactivated to minimize peak tailing caused by interactions with active sites in the sample flow path.

The evaluation includes testing with a specialized Test Mix #1 and linearity studies on various pesticides and semivolatile phenols and amines. Results demonstrate that the HP-5MSi column provides superior peak shapes and linear calibration curves compared to standard columns, allowing for the quantitation of compounds at low nanogram levels.

What's inside

  • Introduction
  • Sample Preparation
  • Chromatographic Conditions
  • Column Evaluation
  • Pesticides Analysis
  • Semivolatile Phenols and Amines
  • Calibration Results
  • Conclusions

Specifications

Column Dimensions30 m x 0.25 mm x 0.25 µm
Carrier Gas (Test Mix #1)He, 1 mL/min
Injector Temperature (Test Mix #1)250 °C
Oven Temperature Range (Test Mix #1)70 °C to 280 °C
Detector Temperature (MSD)280 °C

Frequently asked questions

Which compounds were tested for linearity?

The document tested terbofos, malathion, endrin aldehyde, DDT, 2,4-dinitrophenol, pentachlorophenol, and n-nitrosodipropylamine.

How does the HP-5MSi column perform compared to competitive columns?

It exhibits much improved peak shapes and superior linearity with lower deviation in response factors compared to standard 5ms columns.