Agilent Eclipse Plus Phenyl-Hexyl Column Taxane Analysis Application Note

This application note describes how to use Agilent ZORBAX Rapid Resolution High Throuhput (RRHT) Eclipse Plus Phenyl-Hexyl column for high-resolution analysis of 13 taxane derivatives. Improved silica, endcapping, and bonding technology, as well as the molecular interactions of the phenyl-hexyl moiety with the conjugation within the taxane derivatives, make the Eclipse Plus Phenyl-Hexyl column an ideal choice for this RRHT taxane method.


Agilent Eclipse Plus Phenyl-Hexyl Column Taxane Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
343 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
1EB86CCB35555E460BE2D1969D7CE41C

About this manual

This document details a chromatographic method for separating thirteen taxane derivatives, including paclitaxel and docetaxel, using an Agilent ZORBAX Eclipse Plus Phenyl-Hexyl column. The study demonstrates that this column provides superior resolution compared to C18 columns for these complex diterpene structures.

The method utilizes an Agilent 1200 Rapid Resolution LC (RRLC) system with a gradient elution of 0.1% formic acid in water and acetonitrile. Results show that high resolution is maintained even when a Single Quadrupole Mass Spectrometer is installed in series after the diode array detector, making it suitable for drug discovery and analysis of cancer treatment compounds.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Conclusions
  • References

Specifications

Column Dimensions4.6 x 50 mm
Particle Size1.8 um
Part Number959941-912
Flow Rate1 mL/min
Injection Volume5.0 uL
Column Temperature25 °C
Detection Wavelength254 nm
Maximum Pressureapproximately 240 bar

Frequently asked questions

Which column is recommended for separating taxane derivatives?

The Agilent ZORBAX Eclipse Plus Phenyl-Hexyl column is recommended as it provides excellent separation and better selectivity than C18 columns.

Can the method be used with mass spectrometry?

Yes, no significant loss in chromatographic resolution is noted when an Agilent G1956B Single Quadrupole Mass Spectrometer is installed in series.

What is the maximum pressure reached during the analysis?

The maximum pressure reached during this analysis was approximately 240 bar with the pulse damper used.