Agilent LC/MSD TOF Mass Accuracy Evaluation Application Note

The Application Note describes the application of accurate mass measurement using the ESI TOF technology, and gives mean and standard deviation of repeatedly measured mass accuracies from a real-life pharmaceutical sample.


Agilent LC/MSD TOF Mass Accuracy Evaluation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
239 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
75FA2B2DD4CF5A74032E179895374F1E

About this manual

This document presents a statistical evaluation of mass accuracy measurements performed by an Agilent LC/MSD TOF mass spectrometer. The study utilizes a real-life pharmaceutical sample consisting of degradation products from the antibiotic drug amoxicillin to assess instrument performance under realistic application conditions.

The analysis was conducted using an Agilent 1100 Series capillary LC system coupled with an Agilent LC/MSD TOF equipped with a dual sprayer source. The experiment involved measuring the accurate mass of five identified degradation compounds and their fragments. Statistical analysis of 135 data points yielded an overall mass accuracy of 1.73 ppm with a standard deviation of 0.97 ppm, demonstrating the instrument's capability for reliable empirical formula confirmation.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Methods
  • Sample preparation
  • Results and discussion
  • Statistical evaluation
  • Empirical formula confirmation

Specifications

Overall mass accuracy1.73 ppm
Standard deviation (ppm)0.97 ppm
Confidence interval (3s)99.7 % probability
Mass rangem/z 114 to m/z 515
Instrument resolutionabove 15,000

Frequently asked questions

What is the overall mass accuracy achieved in this evaluation?

The overall mass accuracy was calculated as 1.73 ppm.

What sample was used for the mass accuracy evaluation?

A real-life pharmaceutical sample of degraded antibiotic amoxicillin was used.

Which software was used for data analysis?

Analyst QS software was used for data analysis.