GC/MS Analysis of PCBs in Waste Oil Using the Backflush Capability of the Agilent QuickSwap Accessory

This document presents a practical example of backflushing using the Agilent QuickSwap accessory for analyzing polychlorinated biphenyls (PCBs) in waste oil. The benefits of using QuickSwap instead of the traditional high-temperature bakeout procedure are demonstrated, including significantly reduced column and detector contamination and increased sample throughput.


GC/MS Analysis of PCBs in Waste Oil Using the Backflush Capability of the Agilent QuickSwap Accessory - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
222 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
013D33BD30A0A912CC1CE788C4965560

About this manual

Polychlorinated biphenyls (PCBs) in waste oil are typically analyzed by GC-ECD or GC/MS after solid phase extraction cleanup. However, not all problematic matrix components are completely removed during cleanup and can contaminate the column and detector.

This application presents a practical example of backflushing using the Agilent QuickSwap accessory installed on an Agilent 7890 GC/5975 MSD system. Benefits of using QuickSwap instead of the traditional high-temperature bakeout procedure are demonstrated, showing that column and detector contamination were significantly reduced and sample throughput increased.

What's inside

  • Introduction
  • Sample Preparation
  • GC Conditions
  • Results
  • Backflushing Benefits

Frequently asked questions

What is the primary function of the Agilent QuickSwap accessory?

Its primary function is to simplify changing columns and performing maintenance on GC/MS systems by providing a flow of clean carrier gas that excludes air from the mass spectrometer when columns are disconnected.

How does backflushing remove contaminants compared to traditional bakeout procedures?

Backflushing reverses the column flow so low-volatility contaminants travel only a short distance to be removed, whereas in a traditional bakeout they must travel through the full length of the column.

What is the total cycle time for the oven temperature program that includes backflushing?

The total run time for the oven temperature program with backflush is 22 minutes.