Agilent Technologies G1600 Capillary Electrophoresis Application Note for Heparin Impurities

This Application Note describes a high-speed, high-resolution capillary electrophoresis method of heparin and related impurities using highly concentrated buffer in a 25 µm bubble cell capillary. The method can be used to detect heparin and related impurities in a short time.


Agilent Technologies G1600 Capillary Electrophoresis Application Note for Heparin Impurities - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
246 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
G1600
MD5 checksum
DCD64C033C76B5F377A9EF123E3C77C5

About this manual

This document details an optimized capillary electrophoresis (CE) method for determining impurities in heparin samples, specifically oversulfated chondroitin sulfate (OSCS). The method was developed to improve upon the original USP emergency method by increasing resolution and sensitivity. It utilizes a 25 µm inner diameter bubble cell capillary with highly concentrated lithium phosphate buffers at low pH.

What's inside

  • Abstract
  • Background
  • Method Improvements
  • Experimental design and evolution of the method
  • Operating characteristics of the method
  • Operational parameters

Frequently asked questions

What is the limit of detection for OSCS using this optimized method?

The limit of detection for OSCS is well below 0.1%.

What is the optimal buffer concentration for separating heparin impurities?

The optimal buffer used in this method is 600 mM lithium phosphate at pH 2.5.

What capillary dimensions are required to control the current with high-concentration buffers?

A 25 um inner diameter capillary is required when using such a high buffer concentration.