Agilent 1290 Infinity LC Application Note for Metoclopramide Impurity Analysis

This Application Note evaluates the performance of the Agilent 1290 Infinity LC system for the determination of impurities and related substances of metoclopramide hydrochloride in a pharmaceutical formulation. The translation of a conventional liquid chromatography (LC) method on a 1200 Series HPLC system to an ultra high pressure method on an Agilent 1290 Infinity LC system is discussed. Method translation is relatively easy and temperature fine-tuning is the most important parameter to obtain the same selectivity for the different impurities. Pressures as high as 1070 bar were applied during method development. The final high productivity method is carried out at 880 bar in an analysis time of 3.5 min which is approximately 4 times faster than the original HPLC method but with the same accuracy. A validation study was carried out to demonstrate the performance of the Agilent 1290 Infinity LC system. Limits of detection for the impurities were as low as 0.001 % w/w relative to the main compound using the new diode array detector (DAD). This is more than one order of magnitude lower than required.


Agilent 1290 Infinity LC Application Note for Metoclopramide Impurity Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
516 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
905F9BF285867F6D8765C3DBAC8AF2E7

About this manual

This document evaluates the performance of the Agilent 1290 Infinity LC system for determining impurities and related substances in metoclopramide hydrochloride formulations. It details the translation of a conventional liquid chromatography (LC) method from a 1200 Series HPLC system to an ultra-high pressure method on the 1290 Infinity LC system. The study highlights that method translation is relatively easy, with temperature fine-tuning being the most important parameter to maintain selectivity for different impurities.

The final high productivity method achieves an analysis time of 3.5 minutes, which is approximately four times faster than the original HPLC method, while maintaining the same accuracy. The system utilized a high-pressure pump capable of delivering up to 1200 bar and a new fast and sensitive diode array detector (DAD), achieving limits of detection for impurities as low as 0.001% w/w relative to the main compound.

What's inside

  • Abstract
  • Introduction
  • Experimental Instrumentation
  • Method translation
  • Increasing speed
  • Results and Discussion
  • Final method conditions

Specifications

Maximum pressure1200 bar
Analysis time3.5 min
Detection limit0.001 % w/w
Column particle size1.7-µm
Detector typeDiode Array Detector (DAD)

Frequently asked questions

How much faster is the new method compared to the original?

The final high productivity method is approximately 4 times faster than the original HPLC method.

What is the most important parameter for obtaining the same selectivity?

Temperature fine-tuning is the most important parameter to obtain the same selectivity for the different impurities.

What was the maximum pressure applied during method development?

Pressures as high as 1070 bar were applied during method development.