Analysis of Glyphosate and Aminomethyl Phosphonic Acid by LC/MS Application Note

A liquid chromatography/mass spectrometry method using an electrospray ionization source in positive ion mode was developed for the analysis of glyphosate (N-phosphonomethyl glycine) and its metabolite, aminomethyl phosphonic acid in water. Both glyphosate and its metabolite were derivatized using 9-fluorenyl-methyl chloroformate in buffer solution prior to reversed phase high performance liquid chromatography separation. The method cleanly resolved both target molecules with excellent sensitivity in both positive and negative ion modes.


Analysis of Glyphosate and Aminomethyl Phosphonic Acid by LC/MS Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
159 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
98CF14BD6DE1EFF0476D3D74A2AC399A

About this manual

This document presents a liquid chromatography/mass spectrometry (LC/MS) method developed by Agilent Technologies for the analysis of glyphosate and its metabolite, aminomethyl phosphonic acid (AMPA), in environmental water samples. The method utilizes pre-column derivatization with 9-fluorenylmethyl chloroformate (FMOC) to enhance detection sensitivity.

The procedure details sample preparation steps involving buffer solutions and reagent mixing, followed by analysis using an Agilent 1100 LC/MS system equipped with electrospray ionization. The document provides specific instrument parameters, chromatography conditions, and presents results demonstrating the method's ability to resolve both target molecules in positive and negative ion modes.

What's inside

  • Abstract
  • Background
  • Sample Preparation
  • Instrument Conditions
  • LC Conditions
  • Results
  • Conclusions

Frequently asked questions

What is the minimum sample volume required for this analysis?

The method requires a 100-mL water sample size to achieve adequate sensitivity.

How much of each target molecule can the instrument measure on column?

The instrument can easily measure between 10 and 20 ng on column.