Agilent 6820 GC Analysis of Organochlorine and Pyrethroid Pesticides Application Note

This document describes the use of Agilent 6820 gas chromatograph (GC) with micro-electron capture detector (µECD) to analyze organochlorine and pyrethroid compounds. The sensitivity provided by µECD was much better than the requirements of the routine pesticide residue analysis.


Agilent 6820 GC Analysis of Organochlorine and Pyrethroid Pesticides Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
465 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
6820
MD5 checksum
10F8DEDB3949E8D6F5F047B13B5B71E8

About this manual

The document describes the use of the Agilent 6820 gas chromatograph equipped with a high-sensitivity hECD to analyze organochlorine and pyrethroid compounds. The system was evaluated against Chinese National Standard Method GB/T 5009.146-2003, demonstrating good linearity and sensitivity across wide concentration ranges.

What's inside

  • Introduction
  • Application
  • Experimental Conditions
  • Instrumental Parameters
  • Results: hECD Sensitivity
  • Linear Range and Response Factors

Specifications

Detector TypehECD (high-sensitivity Electron Capture Detector)
Column TypeHP-1, 30 m x 0.32 mm x 0.25 µm
Carrier GasNitrogen, constant head pressure mode
Oven Temperature Program60 °C (1 min), 30 °C/min to 180 °C, 5 °C/min to 260 °C (5 min), 3 °C/min to 280 °C (10 min)
Detector Temperature330 °C
Linear Correlation Coefficient (R²)> 0.995 for all compounds

Frequently asked questions

What pesticides were analyzed in this study?

The study analyzed organochlorine and pyrethroid compounds, including BHCs, heptachlor, aldrin, endosulfan, DDT derivatives, fenpropathrin, cyhalothrin, permethrin, cypermethrin, fenvalerate, and deltamethrin.

Does the hECD meet routine pesticide residue analysis requirements?

Yes, the system shows good sensitivity and a wide linear range that are much better than routine pesticide residue analysis requirements.

Download PDF (465 KB)