Agilent 1100 Series HPLC Application Note: Separation of Antibacterial Drugs

This document discusses the natural occurrence of tetracyclines, their use in human and veterinary medicine, and their use as nutritional antibiotics in pig and poultry farming. The document also demonstrates the HPLC separation of antibacterial drugs with a tetracycline structure using a Zorbax SB-C18 reversed phase column and an alternative mobile phase to TFA. The HPLC method provides an easy, reliable, and precise analysis of the antibacterial drugs.


Agilent 1100 Series HPLC Application Note: Separation of Antibacterial Drugs - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
34 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
66A0E57B4C90DAE7AA87DEE48C0F80E6

About this manual

This document details an HPLC method for separating three common tetracycline analogs: Minocycline, Tetracycline, and Doxycycline. The method utilizes a Zorbax SB-C18 reversed phase column and demonstrates separation without ion pairing, using an alternative mobile phase to trifluoroacetic acid (TFA).

The experimental setup includes an Agilent 1100 Series HPLC with a variable wavelength UV detector set to 350 nm. The method specifies a gradient elution from 5% to 60% acetonitrile in 10 minutes, followed by a column wash. Performance data for the limit of detection (LOD), retention time precision, and area precision are provided for the three compounds.

What's inside

  • Application
  • Experimental Conditions
  • Highlights
  • Equipment
  • HPLC Method Performance

Specifications

ColumnZorbax SB-C18, 3.5 µm, 4.6 x 75 mm
Mobile phase A0.025 M KH2PO4 in water (pH = 3)
Mobile phase BAcetonitrile
Injection volume5 µL
Temperature25 °C
Flow rate1.0 mL/min
Gradient5% B to 60% B in 10 min
Column wash60% B to 5% B in 2 min

Frequently asked questions

Does this method use ion pairing?

No, the method demonstrates separation without ion pairing.

What is the limit of detection (LOD) for Minocycline?

The LOD for Minocycline is 0.1 mg/L.