Agilent Technologies 1100 Series LC-APPI-MS Chloramphenicol Determination Application Note

This article introduces a liquid chromatography-atmospheric pressure photoionization-mass spectrometry method (LC-APPI-MS) for the determination of chloramphenicol antibiotics in fish meat. The method involves extraction of the samples with ethyl acetate, followed by liquid-liquid distribution with acetonitrile and n-hexane for cleanup. Recoveries of chloramphenicol from young yellowtail meat and flatfish meat spiked at 0.1-2 ng/g were 89.3%-102.5% and 87.4%-94.8%, respectively. The limit of detection (signal-to-noise = 3) of the young yellowtail meat and the flatfish meat were 0.27 and 0.10 ng/g.


Agilent Technologies 1100 Series LC-APPI-MS Chloramphenicol Determination Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
99 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
5F636193435CC46950D1B455C7202AD3

About this manual

The document details a method developed for determining chloramphenicol antibiotics in fish meats using liquid chromatography-atmospheric pressure photoionization-mass spectrometry (LC-APPI-MS). The study focuses on optimizing ion source parameters to achieve simple mass spectra and strong signals corresponding to the [M-H] ion.

The experimental procedure involves extracting samples with ethylacetate, followed by a clean-up step using liquid-liquid distribution. Key optimization results include setting the capillary voltage at 3500 V when acetone is used as a dopant, the vaporizer temperature at 350 °C to prevent thermal degradation, and the fragmentor voltage at 120 V for optimal signal-to-noise ratio.

What's inside

  • Abstract
  • Introduction
  • Chemicals and Solvents
  • Sample Preparation
  • LC/MS Instrument Parameters
  • Optimization of APPI Parameters
  • Effect of Capillary Voltage
  • Effect of Vaporizer Temperature
  • Results and Discussion

Specifications

ColumnZorbax Eclipse XDB C18 (150 mm x 3 mm, 5 um)
Ionization ModeAPPI (Negative)
SIM Ion Targetm/z 321; (M-H}
Drying Gas Flow7 L/min at 360 °C
Nebulizer Pressure50 psi
Fragmentor Voltage120V
Capillary Voltage (with dopant)3500 V

Frequently asked questions

What is the optimum capillary voltage for chloramphenicol analysis with acetone as a dopant?

The optimum capillary voltage is 3500 V, which yields higher intensity than without the dopant.

What vaporizer temperature should be used to avoid thermal degradation of chloramphenicol?

The vaporizer temperature should be set at 350 °C, as higher temperatures cause thermal degradation.

What fragmentor voltage provides the best signal-to-noise ratio for chloramphenicol?

The optimum fragmentor voltage is found at 120 V, where the best S/N ratio was observed.