Agilent 7500i ICP-MS Mercury Determination Application Note

This document presents a quantitative determination method for mercury in food using 7500i ICP-MS. Gold is added as a cleansing agent to avoid memory effects. The use of specific instrumental setup and measurement mode allows for rapid and accurate determination of trace mercury levels.


Agilent 7500i ICP-MS Mercury Determination Application Note - cover page
Brand
Agilent
Category
Other products
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
111 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
7500i
MD5 checksum
64CAE67E6D0818D1BF52BC18424EE021

About this manual

This document presents a method for the quantitative determination of mercury in foodstuffs using an Agilent 7500i ICP-MS. The procedure involves preparing microwave digests and analyzing them with gold added to standards, samples, and wash solutions to act as a cleansing agent and prevent memory effects.

The method utilizes the integrated sample introduction system (ISIS) in high sample throughput mode, a second vacuum pump, and a micro-flow concentric nebulizer. This setup allows for robust and rapid determination of mercury at parts per trillion (ppt) levels, achieving excellent agreement with certified reference materials and demonstrating system stability over 36-hour analytical runs.

What's inside

  • Application
  • Introduction
  • Procedure
  • Microwave Digestion
  • Operating and Acquisition Parameters
  • Instrument Conditions
  • ISIS Peristaltic Pump Program
  • Results
  • Quality Control
  • Conclusion

Specifications

Plasma gas flow rate16 L/min
Carrier gas flow rate0.85 L/min
Make-up flow0.14 L/min
RF Power1400 Watts
NebulizerAgilent micro-flow 100 pL
Spray chamber temperatureCooled to 2 °C
Sample tubing0.89 mm id
Internal standard tubing0.19 mm id
Analysis Speed (ISIS)0.10 rps

Frequently asked questions

What instrument is used for mercury determination in this method?

The method uses a 7500i ICP-MS.

How is memory effect avoided during analysis?

Gold is added offline to all standards, samples, and wash solutions to act as a cleansing agent.

What is the typical detection range for mercury in this method?

Detection levels of between 10 and 30 ppt are routinely achieved.

How long does the system demonstrate stability?

The system stability was demonstrated over a 36-hour analytical run.