Agilent Bio-Monolith OA Anion-Exchange HPLC Application Note for Adenovirus Analysis

This document describes a rapid analysis method for adenovirus type 5 particles, using a bio-monolith anion-exchange HPLC column to support the development of a high-titre manufacturing platform. The method incorporates regeneration conditions to extend the column's functional life and efficiently separates intact adenovirus type 5 particles from contaminating proteins and DNA. The method was used to analyze samples throughout a downstream process and the samples were detected at 260 nm and 280 nm using a photodiode array detector (PDA). The method was validated based on performance criteria including repeatability, intermediate precision, and linearity. The linear working range of analysis was established between 7.5 x 10^8 to at least 2.4 x 10^10 viral particles (3 x 10^10 to 9.6 x 10^11 viral particles/mL), with a correlation coefficient of 0.9992. The relative standard deviations (RSDs) for intra- and inter-day repeatability and precision for retention time and peak area were less than 1 and 2.5%, respectively.


Agilent Bio-Monolith OA Anion-Exchange HPLC Application Note for Adenovirus Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
794 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
742080960CAE4AF0C32314D97EFED634

About this manual

The document details the development and validation of an analytical method for quantifying adenoviral particles (Ad5) throughout the biopharmaceutical manufacturing process. The method utilizes strong anion-exchange Bio-Monolith HPLC columns to efficiently separate intact Ad5 particles from contaminating proteins and DNA.

What's inside

  • Abstract
  • Introduction
  • Experimental Conditions
  • Method Development
  • Method Qualification
  • In-process Analysis
  • Conclusions

Frequently asked questions

What is the linear working range for analyzing adenovirus particles?

The method establishes a linear working range between 7.5 x 10^8 and at least 2.4 x 10^10 viral particles.

What flow rate is used for the HPLC analysis?

The assay runs at a constant flow rate of 1 mL/min.

How much sample volume is injected per run?

For each run, 25 µL of adenovirus type 5 was loaded onto the column.