Agilent Fast Sensitive EtG Analysis in Urine With Pursuit C18 HPLC Column the 320-MS LC/MS/MS Application Note

EtG is a minor metabolite of ethanol, can be detected in blood for up to 36 hours and in urine for up to 5 days, can be used as a biomarker for alcohol consumption. This application note describes a fast and sensitive EtG analysis method using the Varian Pursuit C18 HPLC column and 320-MS LC/MS/MS system, which demonstrates good linearity, high sensitivity, and stable ion ratios with a detection limit of 10 ppb.


Agilent Fast Sensitive EtG Analysis in Urine With Pursuit C18 HPLC Column the 320-MS LC/MS/MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
348 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
320-MS
MD5 checksum
2799F963F4851C5E43F01B947E93BC49

About this manual

This document details a method for detecting 8-ethyl glucuronide (EtG), a biomarker for alcohol consumption, in urine samples. The method utilizes the Varian 320-MS LC/MS/MS system combined with Varian Pursuit C18 HPLC columns to achieve high sensitivity and fast separation times.

The analysis demonstrates detection limits of 10 ppb with a reporting limit of 52 ppb. The method involves direct injection of spiked urine samples and utilizes specific ion ratios for quantitation, achieving separation within 2 minutes. The document provides details on instrumentation, materials, HPLC conditions, and results including calibration curves and fragment ion data.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • HPLC Conditions
  • Results & Discussion
  • Discussion
  • Conclusion
  • References

Specifications

Detection Limit10 ppb
Reporting Limit (LOD)52 ppb
Reporting Limit (LOQ)152 ppb
Column TypePursuit C18 5 um, 50 x 4.6 mm ID
Flow Rate0.7 mL/min
Calibration Linearityr² greater than 0.999

Frequently asked questions

What is the detection limit for EtG analysis?

The detection limit of 10 ppb can be easily reached.

What instruments are used for this method?

The method uses a Varian 320-MS LC/MS/MS system with Varian 212-LC Binary Gradient LC/MS Pumps and a Varian Combi PAL AutoSampler.