Agilent 710 ICP-OES Rhodium Determination Application Note

This application note describes a method for determination of rhodium in catalyst oil and aqueous samples using Agilent 710 ICP-OES. The oil catalyst samples were prepared by microwave digestion and the aqueous samples were digested in nitric acid solution. The method provides good sensitivity and reproducibility and is suitable for determination of rhodium concentrations as low as 100 μg/L.


Agilent 710 ICP-OES Rhodium Determination Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
360 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A4A095ACD2372A7654AC3BFBCB426C37

About this manual

The document details a method for analyzing rhodium content in oil catalyst and aqueous samples. It highlights that an axially viewing configuration was selected over radial viewing to achieve better sensitivity, offering up to a 20-fold improvement in detection limits for low expected concentrations.

For the analysis, the Agilent 710 ICP-OES with simultaneous CCD detection was used. Oil catalyst samples required microwave digestion using a Shanghai EU Microwave Chemistry Technology Co. Ltd. model WX-4000 system, while an aqueous sample was measured directly without preparation.

What's inside

  • Introduction
  • Instrumentation
  • Sample Preparation
  • Microwave Digestion Settings
  • Calibration Solutions
  • Results and Discussion

Frequently asked questions

Why was the axially viewing configuration selected for this analysis?

The axially viewing ICP-OES was selected because it provides better sensitivity than the radially viewed configuration, with an improvement in detection limits of up to 20 fold.

How were the calibration solutions prepared for the digested oil samples?

The standard Rh solutions of 40 and 100 mg/L were diluted by a factor of 10 to concentrations of 4 and 10 mg/L respectively.

Download PDF (360 KB)