Agilent Fast GC Analysis of EPA and EU Regulated PAHs Application Note

This application note illustrates the successful use of an Agilent J&W FactorFour VF-17ms column from Agilent Technologies, Inc. for the fast GC/MS analysis of most PAHs listed in the EU and EPA regulations.


Agilent Fast GC Analysis of EPA and EU Regulated PAHs Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
5
File format
PDF
File size
214 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
7862B36110A2E58E47C85B8BC8B92D13

About this manual

The document illustrates a successful method for analyzing 23 priority PAHs listed in European Union regulations, which differ slightly from the 16 compounds covered under US EPA regulations. Due to the toxicity of these compounds and identical mass fragmentation patterns among isomers, effective chromatographic separation is critical.

The analysis utilizes an Agilent J&W FactorFour VF-17ms column with a specific temperature program optimized for speed without compromising resolution. The method achieves baseline separation of difficult peak pairs within approximately 25 minutes using helium as the carrier gas and a quadrupole mass spectrometer detector.

What's inside

  • Introduction
  • Conditions
  • Results
  • Discussion
  • Peak Identification

Frequently asked questions

How long does the fast GC/MS analysis of EPA and EU regulated PAHs take?

The analysis delivers results within 25 minutes without compromising resolution efficiency.

What is the concentration of the sample used in this application note?

The sample concentration for the analysis was set to 5 ppm.

Which column does Agilent recommend for analyzing these PAHs?

Agilent recommends using a J&W FactorFour VF-17ms column for this analysis.