Agilent SpectrAA 400 Determination of Platinum and Palladium Application Note

This document describes a method for the determination of platinum and palladium in chelated systems using GFAAS.


Agilent SpectrAA 400 Determination of Platinum and Palladium Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
120 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
8D220952E185E923875AA6BA70504709

About this manual

This document details the determination of platinum and palladium in chelated systems, specifically cisplatin and related compounds, using Graphite Furnace Atomic Absorption Spectrometry (GFAAS). The study focuses on measuring the stability constants for Pt-DNA and Pd-DNA complexes, utilizing the labile nature of palladium as a model for platinum.

The procedure employs an Agilent SpectrAA 400 Atomic Absorption Spectrometer equipped with a GTA96 Graphite Tube Atomizer. Specific furnace parameters, including ashing and atomization temperatures, are optimized for both elements. The note also compares the use of premixed standards versus the autosampler's automix facility, demonstrating comparable calibration results and reduced operator error.

What's inside

  • Introduction
  • Experimental Method
  • Furnace Parameters
  • Autosampler Parameters
  • Results and Discussion
  • Analysis of Complexes
  • References

Specifications

Instrument ModelAgilent SpectrAA 400 Atomic Absorption Spectrometer
Atomizer ModelGTA96 Graphite Tube Atomizer
Platinum Resonance Line265.0 nm
Platinum Slit Width0.2 nm
Palladium Resonance Line244.8 nm
Palladium Slit Width0.3 nm
Lamp Current5 mA
Platinum Atomization Temperature2700 °C
Palladium Atomization Temperature2600 °C

Frequently asked questions

What instrument is used for the determination of platinum and palladium in this note?

An Agilent SpectrAA 400 Atomic Absorption Spectrometer and GTA96 Graphite Tube Atomizer were employed.

What is the optimal atomization temperature for platinum?

The optimal atomization temperature chosen for platinum is 2700 °C.

Does the autosampler automix facility provide comparable results to premixed standards?

Yes, the automix facility provides a calibration graph that is comparable with that from premixed standards.