Agilent 500 Ion Trap LC/MS with TurboDDS Application Note

This application note presents a method for the analysis of adulterants in weight loss supplements marketed as “all natural.” The analysis is performed using the Agilent 500 Ion Trap LC/MS with Turbo Data Dependent Scanning (TurboDDS), which performs full scan and MS/MS analysis in a single injection. Target adulter- ants were observed and yielded spectra with excellent library matches to the NIST MS Library when spiked into a dietary supplement matrix at 10 µg/mL and then diluted 1:10 in water. A real incurred weight loss product was also analyzed and found to be adulterated with Bumetanide. Unlike other techniques that use targeted analysis to detect known adulterants in a dietary supplement of interest, TurboDDS screens and identifi es both known and unanticipated adulterants in a single experiment.


Agilent 500 Ion Trap LC/MS with TurboDDS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
565 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
212EB1EC2A03F329B44A1375895BBB45

About this manual

This document presents a method for analyzing adulterants in dietary supplements marketed as "all natural" using the Agilent 500 Ion Trap LC/MS with Turbo Data Dependent Scanning (TurboDDS). The analysis targets common adulterants such as loop diuretics, appetite suppressants, and respiratory drugs.

What's inside

  • Abstract
  • Introduction
  • Sample preparation
  • LC/MS analysis
  • Results and Discussion
  • Real sample injection
  • Conclusions
  • References

Specifications

Column2.0 mm x 150 mm Pursuit XRs 3 yp C18
Flow rate0.200 mL/min
Ionization modeESI positive and negative
API drying gas30 psi at 350 °C
Scan range50-500 m/z

Frequently asked questions

What adulterants were targeted in the analysis?

The analysis targeted loop diuretics (Bumetanide and Furosemide), appetite suppressants (Sibutramine, Rimonabant, and NIDA), a respiratory drug (Theophylline), a laxative (Phenolphthalein), and an antiepileptic drug (Phenytoin).

How does TurboDDS differ from targeted analysis?

Unlike targeted analysis which only detects known compounds, TurboDDS performs full scan surveying and MS/MS data collection simultaneously to identify both known and unanticipated adulterants.

What was the concentration of the spiked standards?

The standard samples were made at 10 pg/mL and then diluted 1:10 in water for a final concentration of 1 g/mL.