Agilent 2100 Bioanalyzer Interference of Sybr Green Application Note

This study explored the interference of Sybr®Green I on the performance of Agilent 2100 bioanalyzer. The results showed that Sybr®Green I had little effect on the performance of the analyzer.


Agilent 2100 Bioanalyzer Interference of Sybr Green Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
78 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
96737A18C87919847BB3C3221CE12C42

About this manual

This document presents a study testing whether the presence of Sybr Green I, a common intercalating dye used in real-time PCR, affects the performance of the Agilent 2100 bioanalyzer. The authors investigated if this dye interferes with the Lab-on-a-Chip technology's ability to measure fragment size and purity.

What's inside

  • Abstract
  • Introduction
  • Aim of the study
  • Design of the study
  • Results
  • Discussion

Frequently asked questions

Does Sybr Green I interfere with Agilent 2100 bioanalyzer detection?

The results demonstrate only minor influences of Sybr Green I on the performance of the Agilent 2100 bioanalyzer.

How does Sybr Green I affect amplicon sizing?

Sybr Green I causes a slight decrease in the apparent amplicon size as detected by the DNA 1000 assay.

How does Sybr Green I affect amplicon quantification?

The presence of Sybr Green I leads to an increase in the computed concentration of PCR products.