Agilent 7890N/7000B Rapid Robust Detection of THC Application Note

A robust method for detection of THC and its metabolites in blood has been developed using SPE extraction and GC/MS/MS with backflushing. The dynamic range of quantification was 0.1 to 50 ng/mL for THC and 11-OH-THC, and 1 to 100 ng/mL for THCA, with a run time of 6 minutes and a cycle time of 8 minutes.


Agilent 7890N/7000B Rapid Robust Detection of THC Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
4.4 MB
Published
11 June, 2012
Updated
29 September, 2023
Model
7000B
MD5 checksum
819BC73F2957B0728554AAD1D26976D9

About this manual

The document outlines a rapid and robust analytical method developed by Agilent Technologies for the detection of tetrahydrocannabinol (THC) and its metabolites, 11-OH-THC, and THCA in blood samples. The method utilizes solid-phase extraction (SPE) with High Flow Bond Elut Certify II cartridges followed by derivatization and analysis via Gas Chromatography/Mass Spectrometry (GC/MS).

The procedure employs an Agilent 7890N gas chromatograph equipped with a Low Thermal Mass (LTM) module and backflushing capabilities, coupled to an Agilent 7000B Triple Quadrupole GC/MS system. This setup enables a short run time of 6 minutes and a cycle time of 8 minutes while maintaining high sensitivity and robustness against matrix interferences.

What's inside

  • Abstract
  • Introduction
  • Standards and Reagents
  • Instruments
  • Sample Preparation
  • Analysis Parameters
  • Results
  • Backflushing
  • Low Thermal Mass Modules
  • Dynamic Range

Specifications

Run time6 minutes
Cycle time8 minutes
Dynamic range for THC and 11-OH-THC0.1 to 50 ng/mL
Dynamic range for THCA1 to 100 ng/mL
Correlation coefficient (R²)0.999

Frequently asked questions

What instruments are used in this method?

The experiments were performed on an Agilent 7890N gas chromatograph coupled to a 7000B Triple Quadrupole GC/MS.

What SPE columns are used for extraction?

Agilent High Flow Bond Elut Certify II solid-phase extraction columns were used.

What is the dynamic range for THC quantification?

The dynamic range of quantification was 0.1 to 50 ng/mL for THC.