Agilent 2100 Bioanalyzer RNA Quantitation Comparison Application Note

This Application Note investigates the various techniques for RNA quantitation, including a comparison of the Agilent 2100 bioanalyzer with UV spectrometry and the Ribogreen® quantitation system to access the linear dynamic range, reproducibility, and the effects of contaminants for each quantitation method.


Agilent 2100 Bioanalyzer RNA Quantitation Comparison Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
128 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
32C0205FEB32254D0D718ACF2B604851

About this manual

This document investigates techniques for RNA quantitation, specifically comparing the Agilent 2100 bioanalyzer with the RNA 6000 Nano assay against UV spectrometry and Ribogreen analysis. The study evaluates these methods based on their linear dynamic range, reproducibility, and accuracy in the presence of common contaminants such as genomic DNA and phenol.

The analysis covers the performance of the Agilent 2100 bioanalyzer using the RNA 6000 Nano LabChip kit, which utilizes microfluidic technology and laser excitation. Results indicate that while UV analysis is stable, it is susceptible to contaminants. Ribogreen offers high sensitivity but can be affected by contaminants and phenol. The Agilent 2100 bioanalyzer is highlighted for its ability to accurately quantify RNA in the presence of contaminants while consuming only 1 µL of sample.

What's inside

  • Introduction
  • Materials and Methods
  • Linear dynamic range
  • Reproducibility
  • Influence of sample contaminants
  • Results and Discussion

Specifications

Linear dynamic range25 ng/ul to 500 ng/ul
Linearity R2 (RNA 6000 Nano)0.9976
Linearity R2 (UV analysis)0.9984
Linearity R2 (Ribogreen)0.9977

Frequently asked questions

What is the linear dynamic range of the RNA 6000 Nano assay?

The assay has a specified linear dynamic range of quantitation from 25 ng/ul to 500 ng/ul.

How does the Agilent 2100 bioanalyzer handle contaminants compared to UV analysis?

The 2100 bioanalyzer can accurately quantitate in the presence of contaminants, whereas UV analysis is impacted by contaminants like genomic DNA or phenol.