Agilent Metanephrines in Urine Application Note

Introduces the separation of four metanephrines in urine with an Agilent Polaris C18-A column


Agilent Metanephrines in Urine Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
226 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
527E9FA8B15F4C0FA1AE05989C584850

About this manual

This document details a High-Performance Liquid Chromatography (HPLC) method for separating four specific metanephrines found in urine samples. The method utilizes an Agilent Polaris C18-A column and electrochemical detection (ECD) to analyze both standard and patient urine samples.

The procedure specifies the use of a mobile phase containing 0.1 M KH2PO4, pH 6.0, 20 mg/L 1-octanesulfonic acid, sodium salt, 1 g/L sodium EDTA, and 6% acetonitrile. The analysis is performed at a flow rate of 1.0 mL/min with an injection volume of 10 uL at a temperature of 30°C.

What's inside

  • Introduction
  • Conditions
  • Column
  • Sample
  • Mobile phase
  • Flow rate
  • Temperature
  • Detector
  • Peak identification
  • Figure 1

Specifications

ColumnAgilent Polaris C18-A, 4.6 x 150 mm, 5 um
Mobile phase0.1 M KH2PO4, pH 6.0, 20 mg/L 1-octanesulfonic acid, sodium salt, 1 g/L sodium EDTA, 6% acetonitrile
Flow rate1.0 mL/min
Injection volume10 uL
DetectorECD, Antec Decade, +0.63 V Glassy Carbon — Ag/AgCl

Frequently asked questions

What column is used for the separation?

An Agilent Polaris C18-A column is used for the HPLC separation.

What are the components of the mobile phase?

The mobile phase consists of 0.1 M KH2PO4, pH 6.0, 20 mg/L 1-octanesulfonic acid, sodium salt, 1 g/L sodium EDTA, and 6% acetonitrile.

Which metanephrines are analyzed?

The method analyzes normetanephrine, metanephrine, 3-methoxytyramine, and 4-O-methyldopamine.

Download PDF (226 KB)