Agilent 6460 Triple Quadrupole LC/MS System Application Note

This application note provides a detailed procedure for developing and transferring a bupropion LC/MS/MS method on an Agilent 1290 Infinity LC system and an Agilent 6460 Triple Quadrupole LC/MS system. The developed method is comparable to the original method in terms of sensitivity and precision.


Agilent 6460 Triple Quadrupole LC/MS System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
128 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
E6761EA084EF8012EABC6E1B1A1CA28D

About this manual

This document presents a case study demonstrating the transfer of a bioanalytical LC/MS/MS assay between laboratories using different vendor instruments. Specifically, a method originally developed on a Shimadzu LC 20A coupled to an ABI/SCIEX API 4000 was transferred to an Agilent 1290 Infinity LC System coupled to an Agilent 6460 Triple Quadrupole LC/MS System.

The study evaluated the linearity, precision, and accuracy of drug-spiked plasma samples on the Agilent system. The mobile phase, column, and multiple reaction monitoring (MRM) transitions were kept identical to the original system. The Agilent source parameters were optimized, resulting in comparable sensitivity and successful method transfer with minimal downtime.

What's inside

  • Abstract
  • Introduction
  • Experimental Parameters
  • LC/MS/MS conditions
  • Standard curves
  • Extraction procedure
  • Data acquisition
  • Results and Discussion
  • Conclusions

Specifications

Column temperature40 ± 5 °C
Mobile phase compositionAmmonium formate (10 mM): acetonitrile (20/80, v/v)
Flow rate0.8 mL/min
Run time2 min
Bupropion MRM transition239.8 — 183
Risperidone MRM transition411.1 — 190
Drying gas flow13.0 L/min
Nebulizer pressure40 psig
Dry gas temperature350°C
Capillary voltage3000 V
Sheath gas flow12.0 L/min
Sheath gas temperature350°C
Bupropion collision energy7 eV
Bupropion fragmentor voltage135 V
Risperidone collision energy25 eV
Risperidone fragmentor voltage90 V
Calibration range1 to 500 ng/mL
Minimum R² value> 0.998

Frequently asked questions

What instrument was used for the method transfer?

The method was transferred to an Agilent 6460 Triple Quadrupole LC/MS System coupled to an Agilent 1290 Infinity LC System.

What were the MRM transitions for bupropion and the internal standard?

The MRM transition for bupropion was 239.8 — 183, and for the internal standard risperidone, it was 411.1 — 190.

What was the linearity range of the calibration curve?

The calibration curve was a 10-point curve plotted from 1 to 500 ng/mL using plasma extracted samples.

Did the method transfer require changes to the LC conditions?

No, the mobile phase, column, and MRM settings were kept identical on both systems, only the source parameters were optimized.