Agilent ZORBAX RRHD 300SB-C18 1.8 um Application Note

This application note describes a method for the rapid separation of recombinant human erythropoietin using the Agilent ZORBAX 300SB-C18 1.8 µm Rapid Resolution High Definition (RRHD) column. This method has increased sensitivity, reduced separation time and improved run-to-run reproducibility, which is in great demand by biotechnology and pharmaceutical companies. The C18 StableBond coating technology provides increased sensitivity, exceptional pH, and thermal stability. The use of a 1.8-µm column designed for UHPLC systems significantly reduces analysis time, critical for increasing the efficiency of QC for protein primary structure analysis. The data presented here focus on the fast separation of various recombinant human erythropoietin isoforms. Like many other biomolecules, recombinant EPO protein exhibits heterogeneity due to modifications that occur during manufacturing. Methods are optimized for flow rates, gradient and reproducibility under acidic conditions that contain trifluoroacetic acid.


Agilent ZORBAX RRHD 300SB-C18 1.8 um Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
377 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
8A0F6BDF1D412164C711D60C86503A93

About this manual

This document presents an application note on the fast separation of recombinant human erythropoietin (rEPO) using the Agilent ZORBAX RRHD 300SB-C18 1.8 um column. The study highlights the column's ability to provide increased sensitivity, reduced separation time, and improved run-to-run reproducibility for protein analysis.

What's inside

  • Abstract
  • Introduction
  • Material and Methods
  • HPLC system
  • HPLC column
  • Chemicals and reagents
  • LC methods
  • Results and Discussion
  • Optimization of flow rate
  • Separation of heat-degraded rEPO

Specifications

Column dimensions2.1 x 50 mm
Particle size1.8 um
Maximum pressure1,200 bar
Temperature stability90 °C
Sample concentration1.0 mg/mL
Injection volume3 uL
Flow rate1.0 mL/min
Detector wavelength280 nm

Frequently asked questions

What is the maximum operating pressure of the Agilent 1290 Infinity LC system?

The system operates at a maximum pressure of 1,200 bar.

What is the particle size of the ZORBAX RRHD 300SB-C18 column used?

The column used has a particle size of 1.8 um.

Can the ZORBAX RRHD 300SB-C18 column withstand high temperatures?

Yes, the columns are stable up to 90 °C.