Agilent 1260/1290 LC Analysis of Bisphenol A Leaching from Baby Feeding Bottles Application Note

This Application Note describes a method for analyzing the leaching of bisphenol A from baby feeding bottles. Bisphenol A is an endocrine disruptor that can mimic the body's own hormones and may have negative health effects.


Agilent 1260/1290 LC Analysis of Bisphenol A Leaching from Baby Feeding Bottles Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
423 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
6B440261F6C89F043EA1DC12871DD33F

About this manual

This document presents a method for quantifying bisphenol A (BPA) and bisphenol F (BPF) extracted from baby feeding bottles. The method utilizes an Agilent 1260 Infinity LC system with a ZORBAX Eclipse Plus C18 column and fluorescence detection. It also details a transfer of this method to an Agilent 1290 Infinity LC system using an ultrahigh performance liquid chromatography (UHPLC) column, which demonstrated improved resolution and sensitivity.

What's inside

  • Abstract
  • Introduction
  • Instruments and software
  • Reagents and materials
  • Chromatographic parameters
  • Preparation of standards
  • Sample preparation
  • Procedure
  • Results and Discussion

Specifications

TCC temperature40°C
FLD Excitation230 nm
FLD Emission316 nm
Flow rate0.9 mL/min
LOQ for Bisphenol A1.06 ng/mL
Sample Recovery80%

Frequently asked questions

What is the limit of quantitation for bisphenol A?

The limit of quantitation (LOQ) for bisphenol A was found to be 1.06 ng/mL.

What was the sample recovery rate for bisphenol A?

During sample recovery studies, 80% recovery was obtained for bisphenol A.