Agilent 1260 Infinity Analytical SFC System vs 1200 Series LC System UV Detection Comparison

This application note compares the UV sensitivity and detection limits of the Agilent 1260 Infinity Analytical SFC System and the Agilent 1200 Series LC System. Two applications were analyzed using the same column, detector, and separation conditions for both SFC and RPLC. The results showed that although the Agilent 1200 Series LC System was slightly more sensitive, the Agilent 1260 Infinity Analytical SFC System achieved near-HPLC sensitivity.


Agilent 1260 Infinity Analytical SFC System vs 1200 Series LC System UV Detection Comparison - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
599 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
2BBDD8A5D98C5B3CA217EE22E9A10678

About this manual

This document compares the UV sensitivity and detection limits of the Agilent 1260 Infinity Analytical SFC System against an Agilent 1200 Series LC System. The comparison utilizes two specific applications: a standard mixture of polycyclic aromatic hydrocarbons (PAHs) and the determination of 2,6-dichloroaniline, a potential genotoxic impurity, in diclofenac. To ensure a fair comparison, identical columns, injection volumes, detectors, and flow cells were used for both supercritical fluid chromatography (SFC) and reversed phase liquid chromatography (RPLC) modes.

What's inside

  • Introduction
  • Experimental Solutions
  • System modules
  • Experimental conditions
  • Results and Discussion
  • PAHs
  • Determination of 2,6-dichloroaniline
  • Conclusion

Specifications

Outlet pressure (SFC)120 bar
Flow rate2.0 mL/min
Temperature (SFC)40 °C
Publication Number5990-9195EN

Frequently asked questions

Which system was slightly more sensitive in the comparison?

The Agilent 1200 Series LC System was slightly more sensitive than the SFC system.

What is the detection limit comparison between SFC and LC?

Typically, the limits of detection in LC mode are between two and four times lower than in SFC mode.

What was the limit of quantification for 2,6-dichloroaniline in the API?

The limit of quantification was below the 0.01% impurity level (S/N > 13).