Agilent Bio IEX and MAb Columns Protein Separation Application Note

This application note describes how to use Agilent Bio IEX (WCX and SCX) and MAb nonporous 3 µm poly(styrene divinylbenzene) columns to separate proteins.


Agilent Bio IEX and MAb Columns Protein Separation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
270 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
5E36470A8D4E6D83A24A0C51B01BB30C

About this manual

This document presents a method for protein separation using Agilent Bio IEX (WCX and SCX) and MAb nonporous, 3 µm poly(styrene divinylbenzene) columns. It compares the elution differences between these column types when separating Cytochrome c, ribonuclease A, and lysozyme.

What's inside

  • Introduction
  • Materials and Methods
  • Conditions
  • Results and Discussion
  • Conclusions
  • Table 1

Specifications

Column TypeAgilent Bio IEX (WCX, SCX) and MAb
Particle Size3 µm
Column Dimensions4.6 x 50 mm
Flow Rate0.5 mL/min
TemperatureAmbient
Injection Volume2 µL
Detection Wavelength220 nm
InstrumentAgilent Infinity 1260 Bio-inert HPLC system

Frequently asked questions

What proteins were used in the separation test?

The test used Cytochrome c, ribonuclease A, lysozyme, and a protein mix.

What is the particle size of the columns used?

The columns used have a particle size of 3 µm.

How does the elution order differ between WCX/MAb and SCX columns?

WCX and MAb elute in the order cytochrome c, RNase, then lysozyme, while SCX elutes RNase A, cytochrome c, then lysozyme.