Agilent 7890A GC Analysis of Low-Level Sulfur Compounds Application Note

This document describes the analysis of low-level sulfur compounds in natural gas and propylene using an Agilent 7890A GC system equipped with an OI 5380 pulsed flame photometric detector (PFPD). Reliable routine detection of sulfur components in light hydrocarbon matrices including C3 streams to 50 ppbv is demonstrated.


Agilent 7890A GC Analysis of Low-Level Sulfur Compounds Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
1.2 MB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
8768D4920EF4006209BC3F0A0C105E69

About this manual

The document describes a method for analyzing trace sulfur components, such as hydrogen sulfide and carbonyl sulfide, in light hydrocarbon matrices like natural gas and propylene. The study utilizes an Agilent 7890A Gas Chromatograph equipped with an OI Analytical Model 5380 Pulsed Flame Photometric Detector (PFPD) to achieve reliable detection down to 50 ppbv.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Conclusions
  • Ordering Information

Frequently asked questions

What is the minimum detection limit for sulfur components in light hydrocarbon matrices?

The system can achieve reliable routine detection of sulfur components at levels under 50 ppbv.

What sample loop size and split ratio are required for the initial setup of a PFPD method?

The initial setup must use a 1-mL sample loop and a split ratio of 3 to 1.

How much better is the sensitivity of a tuned PFPD method compared to an FPD?

Once tuned, the sensitivity can be up to 5x better than the FPD.

Download PDF (1.2 MB)

More Agilent Manuals