Agilent ZORBAX RRHD Columns HILIC vs RPLC Morphine Application Note

This application note uses Agilent ZORBAX RRHD columns with UHPLC/MS to compare Morphine’s HILIC and RPLC. Both methods use acetonitrile/ammonium acetate buffer mobile phase for isocratic elution. The results show that HILIC mode allows for more sensitive MS detection due to more efficient spraying and desolvation in the ESI-MS source, as a result of the volatile acetonitrile rich mobile phase used.


Agilent ZORBAX RRHD Columns HILIC vs RPLC Morphine Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
1 MB
Published
08 June, 2012
Updated
29 September, 2023
MD5 checksum
044B4E6795AE3C59017E699B0267FE54

About this manual

This document compares the performance of Hydrophilic Interaction Chromatography (HILIC) and Reversed-Phase Liquid Chromatography (RPLC) for the analysis of morphine and three of its metabolites. The study utilizes Agilent ZORBAX RRHD columns, specifically the HILIC Plus and Eclipse Plus C18, coupled with an Agilent 1290 Infinity LC System and an Agilent 6410A Triple Quadrupole Mass Spectrometer.

The analysis demonstrates that HILIC mode provides more sensitive mass spectrometer detection compared to RPLC for these polar analytes. This improvement is attributed to the use of a highly organic mobile phase in HILIC, which facilitates more efficient spraying and desolvation in the ESI-MS source, resulting in a more intense signal and reduced baseline noise.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Improvements in RPLC/MS Sensitivity
  • Better LC/MS Sensitivity with HILIC Mode

Specifications

Column 1Agilent ZORBAX RRHD HILIC Plus, 2.1 x 100 mm, 1.8 ym
Column 2Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 x 100 mm, 1.8 pm
Column 3Agilent ZORBAX Eclipse Plus C18, 2.1 x 100 mm, 3.5 jim
Column 4Agilent ZORBAX Eclipse Plus C18, 2.1 x 100 mm, 5 pm
Temperature25°C
Flow rate0.4 or 1.0 mL/min
Injection volume2 pL

Frequently asked questions

What is the main advantage of HILIC over RPLC for these analytes?

HILIC allows for more sensitive MS detection due to more efficient spraying and desolvation in the ESI-MS source caused by the volatile acetonitrile-rich mobile phase.

How much does the UHPLC column improve MS sensitivity compared to a traditional column?

The highly efficient sub-2 ym UHPLC column improves MS sensitivity by a factor of 5 compared to a traditional 5 ym column.