LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120 Application Note

This application note describes an analytical method for determination of carbendazim and 3 other fungicide residues in orange juice with easy sample preparation and LC/MS/MS detection. Developed for the 3 most important systemic fungicides, carbendazim, thiabendazole, and imazalil with its metabolite, this method can be easily expanded to include a broader range of pesticides. A simple and rapid extraction procedure provides clean extracts that without any preconcentration allow for detection of analytes at concentrations below 2 ppb. Quantification was performed using triphenyl phosphate as an internal standard. Excellent accuracy (% recovery) and precision (% CV) data were obtained for three concentration levels: 10, 50, and 250 ppb of each analyte.


LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120 Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
693 KB
Published
08 June, 2012
Updated
29 September, 2023
MD5 checksum
8191FC7AF40F5C3BE48B357A946C5B83

About this manual

The document presents a simple extraction procedure using Agilent Bond Elut Plexa cartridges to determine fungicide residues in fruit juice. The method allows for the detection of analytes at concentrations below 2 ppb without preconcentration, utilizing triphenyl phosphate as an internal standard.

Quantification is performed using an Agilent 6460 Triple Quadrupole LC/MS system with an ESI source. The analysis utilizes a superficially porous Agilent Poroshell 120 EC-C18 column to achieve separation within approximately 5 minutes, ensuring excellent sensitivity and low matrix interference.

What's inside

  • Abstract
  • Introduction
  • Analytes
  • Materials and instrumentation
  • Sample preparation
  • LC/MS/MS conditions
  • Results and Discussion

Frequently asked questions

What fungicides and metabolites does this method determine in orange juice?

The method determines carbendazim, thiabendazole, imazalil, and the imazalil metabolite.

What is the maximum pressure the LC system reaches during analysis?

The actual pressure in the LC system does not exceed 400 bar.

What is the recommended centrifugation speed for sample pretreatment?

Samples should be centrifuged for 15-20 minutes at 6000 rpm.