Agilent Synthetic Cannabinoids in Oral Fluid Application Note

This document introduces how to detect synthetic cannabinoids in oral fluid


Agilent Synthetic Cannabinoids in Oral Fluid Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
724 KB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
F7C23A3AA98A35B6C70FC79ABAEEF26B

About this manual

This document details a method for detecting specific synthetic cannabinoid compounds, known as 'Spice' components, in oral fluid specimens. The study utilizes the Quantisal device for non-invasive sample collection and analyzes the specimens using Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

The analytical method employs an Agilent 1200 Series liquid chromatography pump coupled with an Agilent 6430 Triple Quadrupole LC/MS System. The procedure involves solid-phase extraction using Agilent Bond Elut Plexa cartridges followed by analysis via electrospray ionization (ESI) in positive or negative polarity modes. The note provides specific parameters for mobile phases, column conditions, and multiple reaction monitoring (MRM) transitions for various analytes including JWH-018, JWH-073, JWH-200, JWH-250, HU-210, and CP-47,497.

What's inside

  • Introduction
  • Collection devices and reagents
  • Sample preparation
  • Liquid chromatography tandem mass spectrometry
  • Multiple Reaction Monitoring Transitions
  • Recovery from the collection pad
  • Method Validation Data
  • Data Analysis
  • Linearity and sensitivity
  • Matrix effects

Specifications

Limit of Quantitation (LOQ) for JWH-018, JWH-073, JWH-200, CP 474970.5 ng/mL
Limit of Quantitation (LOQ) for CP 47497 C8 and JWH-2502 ng/mL
Column dimensions2.1 x 50 mm, 1.8 um
Column temperature60°C
Injection volume5 pl
Flow rate0.5 mL/min
Nitrogen gas temperature350 °C
Gas flow10 L/min
Nebulizer pressure55 psi

Frequently asked questions

What synthetic cannabinoids were tested in this study?

The study tested JWH-018, JWH-073, JWH-200, JWH-250, HU-210, CP-47,497, and CP-47,497 C8 homologue.

What device was used for oral fluid collection?

The Quantisal device was used for oral fluid collection.

What type of mass spectrometry system was used?

An Agilent 6430 Triple Quadrupole LC/MS System operating in electrospray ionization mode was used.