This document mainly introduces the maintenance and troubleshooting of Agilent CE system, including practical considerations, preventive maintenance, troubleshooting and CE method development and optimization.
This document serves as a quick start guide for the operation and care of Agilent Capillary Electrophoresis (CE) systems. It outlines practical considerations regarding capillaries, samples, buffers, and data analysis to ensure reproducibility in mobility, migration time, and peak area.
The manual provides detailed instructions on preventive maintenance for key components such as the injection system, replenishment system, and vial caps. Additionally, it offers troubleshooting advice and a structured approach for CE method development and optimization.
| Migration time %RSD typical range | 0.3-1% |
|---|---|
| Area %RSD typical range | 13% |
They should be cleaned frequently, for example weekly.
Polyurethane caps with part number 5181-1512 should be used.
The order is: 1. Optimize pH, 2. Increase buffer concentration, 3. Change buffer ion, 4. Add buffer modifier or use coated capillary.