This manual provides an introduction to the Agilent G1978B Multimode Source. It covers the source's components, installation and setup, operation, and maintenance.
This user guide provides instructions for installing, operating, maintaining, and troubleshooting the Agilent G1978B multimode ion source for the 6510 Q-TOF LC/MS system. It includes setup procedures, safety precautions, technical specifications, and guidelines for ionization modes such as ESI, APCI, and mixed modes. The guide also covers daily and weekly cleaning routines, tuning procedures, and method configuration using MassHunter software.
The document is structured into sections covering basic operation, installation steps, reference material including safety symbols and technical specs, and empirical formulas for high-resolution MS calibration.
| Model Number | G1978B |
|---|---|
| Product Type | Multimode Ion Source |
| Compatible System | 6510 Q-TOF LC/MS |
| Manual Part Number | 61978-90030 |
| Edition | First Edition, November 2006 |
In MassHunter software, change the Context to Acquisition, set Ion source to Multimode, choose an ionization mode (ESI, APCI, or Mixed), set source conditions, and save the method.
Never bypass the LC diverter valve; connecting directly to the nebulizer may impair safety protections and cause fire.
The multimode source supports ESI, APCI, and Mixed (simultaneous ESI and APCI) modes.