Agilent Multiple Affinity Removal Spin Cartridge Quick Reference Guide

Use the PROTEIN A column to separate serum proteins, including two wash steps and one purification step.


Agilent Multiple Affinity Removal Spin Cartridge Quick Reference Guide - cover page
Brand
Agilent
Document type
Quick Start Guide
Language
English
Pages
2
File format
PDF
File size
905 KB
Published
07 June, 2012
Updated
30 September, 2023
MD5 checksum
161F5703BBCDB38473083B00C48F6735

About this manual

This guide provides step-by-step instructions for preparing, washing, and eluting samples using the Agilent Multiple Affinity Removal Spin Cartridge. It details dilution, filtration, buffer exchange, and fraction collection procedures.

The guide also includes information on re-equilibration, storage, and analysis of collected fractions, with references to the full instruction guide for more details.

What's inside

  • Dilute and filter sample
  • Prepare spin cartridge
  • Wash and collect flow-through F1
  • Wash and collect flow-through F2
  • Prepare for elution
  • Elute bound fraction
  • Re-equilibration
  • Analyze F1 + F2

Frequently asked questions

What is the part number for the Agilent Multiple Affinity Removal Spin Cartridge?

The part number is 5188-5230.

What buffer is used to dilute the human serum sample?

Buffer A is used to dilute the human serum sample.

What is the centrifuge speed and time for collecting flow-through fraction F1?

Centrifuge at 100 x g for 1.6 min.

What buffer is used to re-equilibrate the cartridge after elution?

Buffer A is used to re-equilibrate the cartridge.