Agilent Multiple Affinity Removal Spin Cartridge Quick Reference Guide

This document describes the method and steps for using magnetic bead affinity chromatography to separate immunoglobulins, including diluting and filtering the sample, preparing the magnetic bead chromatography column, applying the sample, washing and collecting the flow-through F1, washing and collecting the flow-through F2, preparing for elution, eluting the bound fraction, re-equilibrating, and analyzing F1 + F2.


Agilent Multiple Affinity Removal Spin Cartridge Quick Reference Guide - cover page
Brand
Agilent
Document type
Quick Start Guide
Language
English
Pages
2
File format
PDF
File size
1.5 MB
Published
07 June, 2012
Updated
30 September, 2023
MD5 checksum
CA78E099CB8CA7891FBF18D7C640FA13

About this manual

This guide provides step-by-step instructions for preparing, washing, and eluting proteins from the Agilent Multiple Affinity Removal Spin Cartridge using mouse serum. It includes dilution, centrifugation, buffer exchange, and fraction collection steps. The cartridge is designed for proteomics research and can be re-equilibrated for reuse. Additional products and resources are listed for further support.

What's inside

  • Prepare spin cartridge
  • Dilute and filter sample
  • Wash and collect flow-through F1
  • Wash and collect flow-through F2
  • Prepare for elution
  • Elute bound fraction
  • Re-equilibration
  • Analyze F1 + F2

Frequently asked questions

What is the part number for the Agilent Multiple Affinity Removal Spin Cartridge?

The part number is 5188-5289.

What is the sample volume to be diluted for use with this cartridge?

Dilute 25-30-μL mouse serum sample to 200 μL with Buffer A.

What centrifuge speed and time are used to collect flow-through fraction F1?

Centrifuge 1.5 min at 100 x g.

What buffer is used to re-equilibrate the cartridge after elution?

Fill a new 5-mL plastic syringe with 4-mL Buffer A.