Agilent AutoTemp Temperature Calibration Service Manual 70-9037

This document describes the features and characteristics of the AutoTemp Temperature Calibration product.


Agilent AutoTemp Temperature Calibration Service Manual 70-9037 - cover page
Brand
Agilent
Document type
Service Manual
Language
English
Pages
8
File format
PDF
File size
271 KB
Published
07 June, 2012
Updated
29 September, 2023
MD5 checksum
7A9343E2747A02361DEA9294FEB3A090

About this manual

The document details the procedure for calibrating the AutoTemp device, which supports eight temperature channels divided into two sets. The process involves connecting specific standard resistors to designated channel pairs and adjusting internal potentiometers to achieve accurate readings at 70.0 °C and 0.0 °C.

What's inside

  • Limitation of Liability
  • AutoTemp Manifold Channels
  • Two-point Calibration
  • Water Bath Temperature Calibration
  • Temperature Measurement System Operation

Frequently asked questions

What resistors are required to calibrate the AutoTemp channels?

Calibration requires a 394.5 ohm resistor for 70.0 °C, a 1355 ohm resistor for 37.0 °C, and a 7355 ohm resistor for 0.0 °C.

How many temperature channels does the AutoTemp box support?

The unit supports eight temperature channels, with Channels 1 through 4 sharing pots to form one set and Channels 5 through 8 forming a second set.

What is the baud rate setting for the dissolution tester?

The system uses a 9600 baud rate to transmit readings to the dissolution tester.