This document describes the features and characteristics of the AutoTemp Temperature Calibration product.
The document details the procedure for calibrating the AutoTemp device, which supports eight temperature channels divided into two sets. The process involves connecting specific standard resistors to designated channel pairs and adjusting internal potentiometers to achieve accurate readings at 70.0 °C and 0.0 °C.
Calibration requires a 394.5 ohm resistor for 70.0 °C, a 1355 ohm resistor for 37.0 °C, and a 7355 ohm resistor for 0.0 °C.
The unit supports eight temperature channels, with Channels 1 through 4 sharing pots to form one set and Channels 5 through 8 forming a second set.
The system uses a 9600 baud rate to transmit readings to the dissolution tester.