This document is the pre-installation manual for VARIAN Micro-GC. It introduces the installation, use and other information of Micro-GC.
This document serves as a pre-installation guide for all Agilent Micro-GC systems and related products. It outlines essential safety information, including alerts regarding hazardous situations and the use of sealed radioactive sources within the DMD module.
The manual details specific environmental requirements such as temperature ranges (0°C to +50°C) and humidity limits, along with physical space dimensions for various instrument configurations like the 2-CH and 4-CH Micro-GC units. It also specifies power supply needs (12V DC maximum), gas purity standards (99.999%), and minimum computer requirements including a Pentium processor over 1.5 GHz.
| Operating Temperature (Micro-GC) | 0°C to +50°C |
|---|---|
| Humidity Range | 0% to 95% RH (non condensing) |
| Mains Voltage | 100 to 240 Vac, frequency between 50-60 Hz |
| Micro-GC Power Requirement | 12V Vdc, 130 W maximum |
| Gas Purity Minimum | 99.999% |
| Carrier Gas Pressure Range | 200-550kPa +/- 10% (30-80 psi +/- 10%) |
| Minimum Processor Speed | > 1.5 GHz |
No, the instrument is intended strictly for indoor use and must be protected from corrosive chemicals or dust.
The minimum requirement includes Windows XP.