Agilent 240-MS GC/MS Pre-installation Instructions

This document introduces the installation requirements of Varian 240-MS GC/MS Ion Trap Mass Spectrometer, including installation environment, gas supply, power supply, etc.


Agilent 240-MS GC/MS Pre-installation Instructions - cover page
Brand
Agilent
Document type
Installation Guide
Language
English
Pages
14
File format
PDF
File size
509 KB
Published
07 June, 2012
Updated
30 September, 2023
Model
240-MS
MD5 checksum
4D67B2E1DF38A5A2CB34EAFABF6E130A

About this manual

This document provides pre-installation instructions for the Agilent 240-MS GC/MS, covering site preparation requirements such as safety regulations, entrance clearance, bench space, and environmental conditions. It includes specific layout dimensions and load capacities when paired with either the 431-GC or 450-GC models.

What's inside

  • Introduction
  • Checklist Requirements
  • Entrance Clearance
  • Bench Space and Load
  • Vibration Control
  • Temperature and Humidity
  • Particulate Matter

Frequently asked questions

What is the minimum width required for the laboratory entrance?

The entrance to the laboratory must be at least 92 cm (36 in.) wide.

What is the minimum depth required for the bench?

The bench must be at least 84 cm (33 in.) deep.

Download PDF (509 KB)

More Agilent Manuals


View all Agilent manuals →