This instruction manual covers the Sanyo MCO-17AC and MCO-15AC CO2 Incubator. Its contents include precautions for safe operation, cautions for usage, incubator components, installation, operating instructions, alarm and safety functions, routine maintenance, calibration, special functions, troubleshooting, specifications, performance, and a safety check sheet.
The provided text gives safety and usage guidance, including installation by qualified personnel on a sturdy floor, use of a dedicated grounded power source, CO2 gas connection with a pressure regulator, sterile distilled water in the humidifying pan, and closing the inner door completely.
| CO2 sensor type | thermal conductivity type |
|---|---|
| Recommended CO2 gas pressure regulator | 24.5Mpa primary side; 196Kpa secondary side |
| Temperature condition | Chamber temperature at least 5°C higher than ambient |
Only qualified engineers or service personnel should install the unit.
Use sterile distilled water, check the water level every day, and replenish it promptly when low.
The thermal conductivity CO2 sensor is influenced by humidity; lower chamber humidity after door opening can cause a higher-than-actual CO2 display until the humidity recovers.
Fix the unit using the optional kit MCO-17PS.
This manual explains 3 error codes. Open a code for its meaning, causes and what to do: