Error codeE486
The system detects that no instrument is properly connected. The manual's first action is: “Ensure the proper connection of the Olympus HF instrument and / or the connection cable to the UNIVERSAL socket.”
- Model
- Olympus ESG-400
- Display shows
- No instrument connected (E486)
- Type
- Error code
- Source
- Manual p. 126
What E486 means
No instrument connected p. 126
Common causes
- An Olympus HF instrument and / or connection cable has not been properly connected to the UNIVERSAL socket. p. 126
- Malfunction of the Olympus HF instrument and / or the connection cable. p. 126
- Malfunction of the electrosurgical generator. p. 126
What to do
- Ensure the proper connection of the Olympus HF instrument and / or the connection cable to the UNIVERSAL socket. p. 126
- Replace the Olympus HF instrument and / or the connection cable. p. 126
- If the problem persists, contact Olympus. p. 126
Before you service the product, read the safety instructions in the manual p. 9.
If E486 is still not resolved
- If the problem persists, contact Olympus. p. 126 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E486; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.
