Error codeE409
Communication error E409 indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral equipment for correct connection.”
- Model
- Olympus ESG-400
- Display shows
- Communication error (E409)
- Type
- Error code
- Source
- Manual p. 124
What E409 means
Communication error p. 124
Common causes
- Improper connection of the communication cables to the LINK-OUT / LINK-IN socket on the rear panel of the electrosurgical generator or to the peripheral equipment. p. 124
- Malfunction or damage of the communication cables. p. 124
- Malfunction of the electrosurgical generator. p. 124
Safety warnings — read before you start
- DANGERImportant Information – Please Read Before Use. High frequency leakage current or spark discharge may cause user and / or patient burns. p. 9
What to do
- Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral equipment for correct connection. p. 124
- Check the cables for damages and, if necessary, replace the cables. p. 124
- Contact Olympus. p. 124
If E409 is still not resolved
- Contact Olympus. p. 124 “Contact Olympus.”
This page summarises what the manufacturer’s manual says about E409; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.