Error codeE04
The footswitch pedal is pressed while switching on. The manual's first action is: “Release the footswitch pedal.”
- Model
- Olympus ESG-400
- Display shows
- Footswitch pedal pressed (E04)
- Type
- Error code
- Source
- Manual p. 122
What E04 means
Footswitch pedal pressed p. 122
Common causes
- The footswitch pedal is pressed while switching on. p. 122
- Malfunction of the footswitch. p. 122
- Malfunction of the electrosurgical generator. p. 122
What to do
- Release the footswitch pedal. p. 122
- Replace the footswitch. p. 122
- If the problem persists, contact Olympus. p. 122
Before you service the product, read the safety instructions in the manual p. 9.
If E04 is still not resolved
- If the problem persists, contact Olympus. p. 122 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E04; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.
